How does increasing voltage change an ionization chamber into a proportional counter?
An ionization chamber functions as a proportional counter when the applied voltage is increased to a level that creates an intense electric field near the axial wire.1
In a standard ionization chamber, the electric field is only strong enough to collect the ion pairs created by incident radiation. By increasing the voltage, the field strength becomes sufficient to accelerate electrons to energies high enough to cause further ionization through collisions with gas molecules. This process, known as gas multiplication, triggers a Townsend avalanche, where a single free electron produces a shower of many additional electrons.213
Because each original electron creates its own individual avalanche, the total charge collected remains proportional to the initial ionization produced by the radiation. This multiplication increases the output pulse size by factors of hundreds or thousands, allowing the device to differentiate between particles of various energies and types, which is the defining characteristic of a proportional counter.31