How do electric and magnetic fields alter the paths of ions in a mass spectrometer?

Created from Britannica articles using AI. AI can make mistakes, so verify using Britannica articles.

In a mass spectrometer, electric and magnetic fields are used to manipulate the paths of ions to separate them based on their mass-to-charge ratios.1

Electric Fields

Electric fields are primarily used to accelerate ions from the source into the mass analyzer using a system of electrostatic slits. Additionally, in devices like the quadrupole mass spectrometer, superimposed radio-frequency and constant potentials are applied between parallel rods. This creates an electric field that acts as a mass filter, allowing only ions within a specific mass range to maintain stable oscillations and reach the detector.12

Magnetic Fields

When ions move through a magnetic field perpendicular to their direction of travel, they follow a circular path. For a monoenergetic ion beam, the radius of this path depends on the ion's mass-to-charge ratio (m/z). The relationship is defined by the equation:3
mz=B2r22V
where B is the magnetic field strength, r is the radius of curvature, and V is the accelerating voltage. By controlling these parameters, the magnetic field effectively separates ions into their various mass components.3